During microwave pressure digestion, risks generally arise as a result of reactions going "out of control". The innovative sensor, control, and safety concept of the speedwave®four reduces these risks to a minimum, providing the highest possible level of safety in all operating conditions.
- Durable, pressure-resistant PFA-coated stainless steel housing;
- Swiveling lid with electronic lock;
- 3 interlock switches to monitor the swing top lid;
- 3 excess temperature switches to monitor the oven chamber and the electronics;
- Temperature and pressure monitoring of all samples in real time;
- Shut down in case of critical operating conditions;
- Reliable rupture discs to provide protection in case of excess pressure;
- Integrated gas collection system to effectively prevent emissions;
- Controlled release of residual pressure when the digestion vessels are opened.